Cleaning
     The pipette must be cleaned, as described below,
     before it is decontaminated. soap solution is
     recommended for cleaning PIPETMAN Classic.
     EXTERNAL
     1.   Remove the tip ejector.
     2.   Wipe the tip ejector with a soft-cloth or
         lint-free tissue impregnated with soap   MAINTENANCE
         solution.
     3.   Wipe the entire pipette with a soft-cloth
         or lint- free tissue impregnated with soap
         solution, to remove all dirty marks. If the
         pipette is very dirty, a brush with soft plastic
         bristles may be used.
     4.   Wipe the entire pipette and the tip ejector
         with a soft cloth or lint-free tissue soaked
         with distilled water.
     5.   Refit the tip ejector and allow the pipette to
         dry.
     INTERNAL
     The following components only can be immersed
     in a cleaning solution: connecting nut, tip ejector,
     tip holder, piston assembly, seal and O-ring.
     1.   Disassemble the pipette as described in
         MAINTENANCE on page 15.
     2.   set aside the upper part in a clean, dry
         place.
     3.   Clean the individual components of the
         lower part of the pipette using an ultrasonic
         bath (20 minutes at 50°C) or with a
         soft-cloth and brushes. small round brushes
         with soft plastic bristles may be used to
         clean the interior of the tip holder.
     4.   Rinse the individual components with
         distilled water.
     5.   leave the parts to dry by evaporation or
         wipe them with a clean soft-cloth or lint-free
         tissue.
     6.   Reassemble the pipette as described in
         MAINTENANCE on page 15.









 MAINTENANCE | PIPETMAN® ClAssIC  PIPETMAN® ClAssIC | UsER’s GUIDE  19