Leak Test
           This test may be performed at any time to check
           that the pipette does not leak, especially after
           performing a maintenance or decontamination
           procedure. If a pipette fails this test, replace
           the O-ring and seal. After making sure that the
      TROUBLESHOOTING  P2 to P200 Models
           pipette is correctly reassembled, repeat this test.

              Fit a PIPETMAN® DIAMOND tip.
           1.
              set the pipette to the maximum volume,
           2.
              given in the specifications, and pre-rinse.
              Aspirate the set volume from a beaker of
           3.
              distilled water.
              Maintain the pipette in the vertical position
           4.
              and wait for 20 seconds.
           5.   If a water droplet appears at the end of the
              tip, there is a leak.
           6.   If you see no droplet, re-immerse the tip
              below the surface of water.
           7.   The water level inside the tip should remain
              constant; if the level goes down, there is a
              leak.
           P1000 to P10mL Models
           1.   Fit a PIPETMAN® DIAMOND tip.
           2.   set the pipette to the maximum volume
              given in the specifications.
           3.   Aspirate the set volume from a beaker of
              distilled water.
           4.   Maintain the pipette in the vertical position
              and wait for 20 seconds.
           5.   If a water droplet appears at the end of the
              tip, there is a leak.