Page 9 - Instruction Manual 2
P. 9

Dispense
     1.   Place the end of the tip against the inside
         wall of the recipient vessel (at an angle of 10°
         to 40°).
     2.   Press the push button slowly and smoothly
         to the first stop.
     3.   Wait for at least a second, then press the
         push button to the second stop to expel   PIPETTING
         any residual liquid from the tip. Keep the
         push button pressed fully down and (while
         removing the pipette) draw the tip along the
         inside surface of the vessel.
     4.   Release the push button smoothly. Eject the
         tip by pressing firmly on the tip of the ejector
         button.
     Guidelines for Good Pipetting

                IMMERsION DEPTH   WAIT TIME
       MODEl
                 (MIllIMETERs)   (sECONDs)
         P2           1              1
         P10          1              1
         P20          2–3            1
        P100         2–4             1
        P200         2–4             1
        P1000        2–4            2–3
       P5000         3–6            4–5
        P10mL         5–7           4–5
     1.   Make sure that you operate the push button
         slowly and smoothly.
     2.   When aspirating, keep the tip at a constant
         depth below the surface of the liquid (refer
         to the table).
     3.   Change the tip before aspirating a different
         liquid, sample, or reagent.
     4.   Change the tip if a droplet remains at the
         end of the tip from the previous pipetting
         operation.
     5.   Each new tip should be pre-rinsed with the
         liquid to be pipetted.








 PIPETTING | PIPETMAN® ClAssIC  PIPETMAN® ClAssIC | UsER’s GUIDE  9