Pre-Rinse the Tips
           some liquids (e.g., protein-containing solutions
           and organic solvents) can leave a film of liquid
           on the inside wall of the tip; pre-rinse the tip to
           minimize any errors that may be related to this
           phenomenon.
      PIPETTING  Pre-rinsing consists of aspirating the first
           volume of liquid and then dispensing it back to
           waste. subsequent volumes that you pipette
           will have levels of accuracy and precision within
           specifications.

           Aspirate
           1.   Press the push button to the first stop (this
              corresponds to the set volume of liquid).
           2.   Hold the pipette vertically and
              immerse the tip in the liquid (refer to
              Guidelines for Good Pipetting on page 9).
              Release the push button slowly and smoothly
              (to top position) to aspirate the set volume of
              liquid. Wait one second, the time depends on
              the model (refer to the table), then withdraw
              the pipette tip from the liquid. You may wipe
              any droplets away from the outside of the tip
              using a medical wipe; however, if you do so,
              take care to avoid touching the tip’s orifice.


             First stop
          Second stop

















           Figure 6
           Pipetting Motion - Aspirate and Dispense