sYMPTOM    POssIBlE CAUsE    ACTION
                  Improper repair or   Refer to
                     assembly       page 15
       Pipette is                   Tighten
       inaccurate  Unscrew tip holder  connecting-nut
                  Connecting nut is   Tighten
                      loose       connecting-nut
                                    Tighten
                  Tip holder is loose  connecting nut  TROUBLESHOOTING
                  Connecting nut is   Tighten
                      loose       connecting nut
                  Incorrect operator
       Pipette is    technique   Operator training
       not precise  Damaged or   Return pipette to
                  corroded piston   supplier
                                  Replace the tip
                 Damaged tip holder  holder
                                  Replace both
                 Worn O-ring or seal  parts

                                  Use PIPETMAN
                   Low quality tips  DIAMOND tips
                                  Replace the tip
                 Damaged tip holder  holder
                                   Replace tip
       Tips fall off   Damaged tip ejector  ejector
       or do not fit   Ejector spacer is   Mount the spacer
        correctly    missing     on the tip ejector
                 The ejector spacer is   Replace the
                     damaged      ejector spacer
                  The tip ejector is   Assemble the tip
                      loose      ejector properly
                  The ejector lock is   Align the ejector
                    misaligned        lock
     However, if you can’t solve the problem, contact
     your Gilson representative.