sYMPTOM POssIBlE CAUsE ACTION
Improper repair or Refer to
assembly page 15
Pipette is Tighten
inaccurate Unscrew tip holder connecting-nut
Connecting nut is Tighten
loose connecting-nut
Tighten
Tip holder is loose connecting nut TROUBLESHOOTING
Connecting nut is Tighten
loose connecting nut
Incorrect operator
Pipette is technique Operator training
not precise Damaged or Return pipette to
corroded piston supplier
Replace the tip
Damaged tip holder holder
Replace both
Worn O-ring or seal parts
Use PIPETMAN
Low quality tips DIAMOND tips
Replace the tip
Damaged tip holder holder
Replace tip
Tips fall off Damaged tip ejector ejector
or do not fit Ejector spacer is Mount the spacer
correctly missing on the tip ejector
The ejector spacer is Replace the
damaged ejector spacer
The tip ejector is Assemble the tip
loose ejector properly
The ejector lock is Align the ejector
misaligned lock
However, if you can’t solve the problem, contact
your Gilson representative.