Chapter 2

            SETTING THE VOLUME

           The volume of liquid to be
      SETTING THE VOLUME  red to indicate the
           aspirated is set using the
           volumeter. The dials are
           colored either black or
           position of the decimal
           point, depending on the
           model (see examples).
           The volume is set by
           turning the thumbwheel
           or the push button. The
           push button makes it
           easier and quicker to
           set volumes, especially
           when wearing gloves.
           The thumbwheel may be
           turned to slowly reach the
           required setting.      Figure 3
           To obtain maximum      Dial Colors by Model
           accuracy when setting the
           volume:                 Push button
              ●  When decreasing the    Thumbwheel
              volume setting, slowly
              reach the required
              setting, making
              sure not to pass the
              setting.
              ●  When increasing the
              volume setting, pass
              the required value by
              1/3 of a turn and then
              slowly decrease to reach   Figure 4
              the volume, making sure   location of
              not to pass the setting.  Push Button and
                                    Thumbwheel
                          COlOR OF VOlUMETER NUMBERs
              MODEl
                         BlACK   RED    INCREMENT
                P2        µL    0.01 µL  0.002 µL
             P10 to P20   µL    0.1 µL   0.02 µL
             P100–P200    µL      —       0.2 µL
            P1000–P5000  0.01 mL  mL     0.002 mL
               P10mL      mL    0.1 mL   0.02 mL

     6   sETTING THE VOlUME | PIPETMAN® ClAssIC                       PIPETMAN® ClAssIC | UsER’s GUIDE