Maximum Permissible Errors
The data provided in the following tables
conform to the IsO 8655-2 standard. With
a precise pipetting technique (refer to
Guidelines for Good Pipetting on page 9) the
P2 model may be used to aspirate volumes as low
as 0.1µl and the P10 model as low as 0.5 µl.
GIlsON IsO 8655
VOlUME* INTRODUCTION
sYsTEMATIC RANDOM sYsTEMATIC RANDOM
ERROR ERROR ERROR ERROR
P2 WITH D10 OR Dl10 TIPs (P/N F144801)
Min 0.2 ±0.024 ≤0.012 ±0.08 ≤0.04
0.5 ±0.025 ≤0.012 ±0.08 ≤0.04
Max 2 ±0.030 ≤0.014 ±0.08 ≤0.04
P10 WITH D10 OR Dl10 TIPs (P/N F144802)
Min 1 ±0.025 ≤0.012 ±0.12 ≤0.08
5 ±0.075 ≤0.030 ±0.12 ≤0.08
Max 10 ±0.100 ≤0.040 ±0.12 ≤0.08
P20 WITH D200 TIPs (P/N F123600)
Min 2 ±0.10 ≤0.03 ±0.20 ≤0.10
5 ±0.10 ≤0.04 ±0.20 ≤0.10
10 ±0.10 ≤0.05 ±0.20 ≤0.10
Max 20 ±0.20 ≤0.06 ±0.20 ≤0.10
P100 WITH D200 TIPs (P/N F123615)
Min 20 ±0.35 ≤0.10 ±0.80 ≤0.30
50 ±0.40 ≤0.12 ±0.80 ≤0.30
Max 100 ±0.80 ≤0.15 ±0.80 ≤0.30
P200 WITH D200 OR D300 TIPs (P/N F123601)
Min 50 ±0.50 ≤0.20 ±1.60 ≤0.60
100 ±0.80 ≤0.25 ±1.60 ≤0.60
Max 200 ±1.60 ≤0.30 ±1.60 ≤0.60
P1000 WITH D1000 OR D1200 TIPs (P/N F123602)
Min 200 ±3 ≤0.6 ±8 ≤3.0
500 ±4 ≤1.0 ±8 ≤3.0
Max 1000 ±8 ≤1.5 ±8 ≤3.0
P5000 WITH D5000 TIPs (P/N F123603)
Min 1000 ±12 ≤3 ±40 ≤15
2000 ±12 ≤5 ±40 ≤15
Max 5000 ±30 ≤8 ±40 ≤15
P10Ml WITH D10Ml TIPs (P/N F161201)
Min 1 ±30 ≤6 ±60 ≤30
2 ±30 ≤6 ±60 ≤30
5 ±40 ≤10 ±60 ≤30
Max 10 ±60 ≤16 ±60 ≤30
*All Values in Microliters